The ADAMS A1c, HA-8180T is a fully automated Glycohemoglobin analyzer based on HPLC (High Performance Liquid Chromatography) and is highly effective for ß-thalassemia testing.
Samples | Whole blood or hemolysis sample |
---|---|
Measurement items | HbA1c(Stable HbA1c, S-A1c), HbA2 and HbF (HbS, HbC, HbE and HbD can be detected.) |
Reagents *1 | ELUENT 80A, ELUENT 80B, ELUENT 80CT and HEMOLYSIS WASHING SOLUTION 80H |
Column *2 | COLUMN UNIT 80T |
Measurement principle | Reversed-phase cation exchange chromatography |
Detection method | Dual-wavelength colorimetry (wavelengths measured: 420nm/500nm) |
Resolution | 0.1% Ratio, 1 mmol/mol |
Measurement range *3 | HbA1c : 3-20%, 9-195 mmol/mol HbA2 : 2-10% HbF : 0-100% |
Guaranteed measurement range *4 |
HbA1c : 4.2-12.9%, 22-117 mmol/mol
HbA2 : 2.4-6.3% HbF : 0.1-14.3% |
Processing speed | 210 seconds per sample |
Sample consumption | Approx.14μL(whole blood) |
Sample container | Blood collection tube (12.3 to 15mm diameter) × ( 75 to 100mm length) Sample cup(500μL) |
Sampling method | Cap-Piercing |
Rack type | ARKRAY racks |
Number of measurement samples | One-way transportation (factory set) : Maximum 50 samples Loop transportation : Maximum 100 samples |
Column temperature | Approx. 39℃ |
Display | Color graphic LCD with backlight |
Memory capacity | 500 test results ( including calibration results ) |
Built-in Printer | Termal printer, 58mm thermal paper |
External output | Serial 1 port (Can be optionally used as an Ethernet port.) |
Measurement conditions | Temperature : 15-30℃ Humidity : 20-80% relative humidity (non-condensing ) |
Required sample volume | Blood collection tube : Minimum 10mm away from the bottom Sample cup : 400μL or more |
Warm-up time | Maximum 30 min. |
Dimensions | 530(W)×530(D)×530(H)mm (Not including the hemolysis washing solution bottle) |
Weight | Analyzer: Approx. 39kg, Sampler: Approx. 4kg |
Power supply | AC100V-240V ±10% 50/60Hz |
Power consumption | Max. 300 VA |
*1, *2 Reagents and column are sold separately from HA-8180T analyzer.
*3 Error-free measurement ranges.
*4 Guaranteed ranges for obtaining results equivalent to those from HA-8160 TP mode measurement.
※Design and specications may be changed without prior notice
AG211206-01A
We use (i) Necessary cookies to provide you with services on the website, (ii) Functional cookies to ensure the proper functioning of browsing, (iii) Analytical cookies to collect information to analyze the traffic to the website and (iv) Performance cookies to count visits and traffic sources in order to improve website performance.
You can inhibit Functional cookies and/or both Analytical and Performance cookies by clicking on OPTIONS.
By clicking ACCEPT, it will be deemed that you accept all the mentioned types of cookies.
When you visit our website, we use the following cookies for a variety of purposes and to enhance your online experience on our website. You can change your preferences and accept/decline certain types of cookies that would otherwise be stored on your computer while browsing our website by adding/removing check marks. But keep in mind that deleting cookies may have an impact on your experience with our website.